Parametro | Valore |
Scala pH |
6.5 to 8.5 pH |
Risoluzione pH |
0.1 pH |
Accuratezza pH |
±0.1 pH of reading at 25°C |
Metodo pH |
Adaptation of the Phenol Red Method |
Scala Bromo |
0.00 to 10.00 mg/L (as Br2) |
Risoluzione Bromo |
0.01 mg/L |
Accuratezza Bromo |
±0.08 mg/L ±3% of reading at 25°C |
Metodo Bromo |
adaptation of Standard Methods for the Examination of Water and Wastewater, 23rd Edition, DPD Method |
Scala Cloro Libero |
0.00 to 5.00 mg/L (as Cl2) |
Risoluzione Cloro libero |
0.01 mg/L |
Accuratezza Cloro Libero |
±0.03 mg/L ±3% of reading at 25 °C |
Scala Cloro Totale |
0.00 to 5.00 mg/L (as Cl2) |
Risoluzione Cloro Totale |
0.01 mg/L |
Accuratezza Cloro Totale |
±0.03 mg/L ±3% of reading at 25°C |
Metodo Cloro |
adaptation of the US EPA Method 330.5, DPD Colorimetric Method |
Scala Acido Cianurico |
0 to 80 mg/L (as CYA) |
Risoluzione Acido Cianurico |
1 mg/L |
Accuratezza Acido Cianurico |
±1 mg/L ±15% of reading at 25 °C |
Metodo Acido Cianurico |
adaptation of the Turbidimetric Method |
Scala Iodio |
0.0 to 12.5 mg/L (as I2) |
Risoluzione Iodio |
0.1 mg/L |
Accuratezza Iodio |
±0.1 mg/L ±5% of reading at 25°C |
Metodo Iodio |
adaptation of Standard Methods for the Examination of Water and Wastewater, 23rd Edition, DPD Method |
Sorgente Luminosa Fotometro/Colorimetro |
LED with 525 nm bandpass filter |
Rilevatore di Luce Fotometro/Colorimetro |
silicon photocell |
Larghezza Filtro Banda |
8 nm |
Accuratezza Lunghezza d'Onda Filtro Banda |
±1.0 nm |
Tipo di Cuvetta |
round 24.6 mm diameter (22 mm inside) |
GLP |
yes |
Display |
128 x 64 pixel B/W LCD with backlight |
Memoria di Registrazione |
50 readings |
Tipo/ Durata Batteria |
1.5 V AA Alkaline (3 pcs.) / > 800 measurements (without backlight) |
Spegnimento Automatico |
after 15 minutes of inactivity (30 minutes before a READ measurement) |
Condizioni di Utilizzo |
0 to 50°C (32 to 122°F); 0 to 100% RH (IP67) |
Peso |
380 g (13.4 oz.) |
Dimensioni |
142.5 x 102.5 x 50.5 mm (5.6 x 4.0 x 2.0") |
Informazioni Ordine |
HI97101C is supplied with CAL Check standards, cuvettes (2), caps (2), plastic inserts for cuvette, (2), scissors, cuvette cleaning cloth, 1.5V AA batteries (3), instruction manual, and instrument quality certificate packaged in a rugged, thermoformed, carrying case. |
HI97101 is supplied with cuvettes (2), caps (2), plastic inserts for cuvettes (2), 1.5V AA batteries (3), instruction manual, and instrument quality certificate. |